Calibration Procedures Teledyne API Model T360/T360M Operation Manual
178
5.5. MANUAL CALIBRATION CHECKS WITH ZERO/SPAN
VALVES
Zero and span checks using the various zero/span valve options available for the
T360 are similar to that described in Section 5.3, except that the zero and
calibration operations are initiated directly
and independently with dedicated
buttons CALZ and CALS.
To perform a manual calibration check of an analyzer with a valve option
installed, use the following method.
STEP ONE: Connect the sources of Zero Air and Span Gas as shown in Figure
5-3 through Figure 5-6.
STEP TWO: Perform the zero/span check.
Wait until STABIL
falls below 1.0
ppm.
This may take
several minutes.
The val ue of STA BIL
may jump
signi ficantly. Wait
until STABIL falls
below 1.0 ppm. This
may take several
minutes.
SPAN CAL M STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > ZERO SPAN CONC EXIT
SAMPLE ST ABIL=XXX.X PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SET UP
ZERO CAL M STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > ZERO CONC EXIT
SAMPLE STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
EXIT returns to the main
SAM PL
E
dis
p
la
y
ACTION:
Record the
CO2 readings
presented in the
upper right c orner of
the display.
ACTI O N:
Rec ord the
CO
2
readings
presented in the
upper r ight corner of
the display.
SAMPLE* RANGE = 500.000PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
Scroll to the STABIL test
function.
07272B DCN6552