Teledyne API – T101 Operation Manual Calibration Procedures
171
Step Two: Set the expected H
2
S span gas value:
This seque nce ca uses the
analyze r to prompt for the
exp ected H
2
S span
con cen tr ation.
M-P CAL H2S SPAN CONC:
50.0 Conc
0 0 0 4 5 0 .0 ENTR EXIT
M-P CAL RANGE = 500.000 PPB H2S =XXX.X
< TST TST > ZERO CONC EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX. X
< TST TST > CAL SE TUP
The H
2
S sp an co nce ntration
valu es a utomati cally defaul t
to 450.0 p pb.
To change this value to th e
actua l concent ra ti on o f the
sp an g as, enter the numbe r
by pressing the key under
each digit until the expected
val ue appear s.
The span gas concentr ation
should always be 90% of the
se lected reportin g rang e
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
EXIT ign ores the new set tin g
and retur ns to the pr evious
display.
ENTR accepts the n ew setting
and returns to the
pre vious disp lay..