MODEL T100H UV FLUORESCENCE SO2 ANALYZER Addendum to T100 Operation Manual, PN 06807 © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CA
TABLE OF CONTENTS Teledyne API - T100H Addendum to T100 Operation Manual x 10.3.2. Sensor Module Repair & Cleaning...
11 1. INTRODUCTION This addendum is based on the Model T100 Operators Manual (P/N 06807). In most ways the T100H analyzer is identical to the T1
Introduction Teledyne API - T100H Addendum to T100 Operation Manual) 12 Operation Instructions This section includes step by step instructions for
13 2. SPECIFICATIONS, APPROVALS & WARRANTY 2.1. Specifications Table 2-1: Model T100H Basic Unit Specifications Min/Max Range (Physical
Specifications, Approvals & Warranty Teledyne API - T100H Addendum to T100 Operation Manual) 14 2.2. CE Mark Compliance Emissions Compliance T
15 3. GETTING STARTED 3.1. Unpacking and Initial Setup CAUTION To avoid personal injury, always use two persons to lift and carry the Model T10
Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 16 3.1.1. Electrical Connections: For full details on the electrical conne
Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 17 3.2. Pneumatic Connections NOTE To prevent dust from getting into the ana
Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 18 3.2.1.2. Connections with Internal Valve Options Installed There is
Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 19 3.2.2. T100H Layout FLOW CONTROL & PRESSURE SENSOR VACUUM PRESSURE SE
Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 20 Figure 3-4: T100H Layout (Basic Unit – No Valve Options) 07265A DC
Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 21 3.3. Initial Operation With the following exceptions, the operation of th
Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 22 3.3.3. Interferents for SO2 Measurements Hydrocarbons are a significan
23 4. FREQUENTLY ASKED QUESTIONS (FAQs) More FAQs are included in the T100 manual, which also includes a glossary of terms. Q: How long does th
Frequently Asked Questions (FAQs) Teledyne API - T100H Addendum to T100 Operation Manual) 24 This page intentionally left blank.07265A DCN6038
25 5. OPTIONAL HARDWARE AND SOFTWARE With the following additions, changes and exceptions, the options listed in Table 1-1 of the T100 Manual (P
Optional Hardware and Software Teledyne API - T100H Addendum to T100 Operation Manual) 26 5.2. Ambient Zero / Two Ambient Span Valve Option (OPT 5
Teledyne API - T100H Addendum to T100 Operation Manual Optional Hardware and Software 27 Table 5-1 describes the state of each valve during the an
Optional Hardware and Software Teledyne API - T100H Addendum to T100 Operation Manual) 28 between the inner and outer tubes is small and the scrub
29 6. OPERATING INSTRUCTIONS 6.1. Warning Messages Please refer to the T100 User’s Manual (06807), Section 4.2.2, for a list of warnings for the
iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage.
Operating Instructions Teledyne API - T100H Addendum to T100 Operation Manual) 30 6.2.2. Range Units The T100H only displays concentrations in pa
Teledyne API - T100H Addendum to T100 Operation Manual Operating Instructions 31 6.2.4. Default DAS Channels The default Data Channels included in
Operating Instructions Teledyne API - T100H Addendum to T100 Operation Manual) 32 6.2.5.2. Control Inputs ZERO CALHI SPAN CALZERO CALCONTROL IN
33 7. CALIBRATION AND CALIBRATION CHECK PROCEDURES Calibration procedures for the T100H are the same as those for the T100. One exception to t
Calibration and Calibration Check Procedures Teledyne API - T100H Addendum to T100 Operation Manual) 34 STEP THREE: Perform the calibration accor
Teledyne API - T100H Addendum to T100 Operation Manual Calibration and Calibration Check Procedures 35 7.2. Manual Calibration Check with Ambient
Calibration and Calibration Check Procedures Teledyne API - T100H Addendum to T100 Operation Manual) 36 This page intentionally left blank.07265A
37 8. INSTRUMENT MAINTENANCE 8.1. Maintenance Schedule There is no Internal IZS offered for the T100H. 8.2. Predictive Diagnostics Because the T
Instrument Maintenance Teledyne API - T100H Addendum to T100 Operation Manual) 38 This page intentionally left blank.07265A DCN6038
39 9. THEORY OF OPERATION 9.1. The UV Light Path The UV light path of the T100H is similar to that of the T100 (see Section 10.2 of the T100 Man
Teledyne API - T100H Addendum to T100 Operation Manual iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long
Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 40 9.1.1. The Reference Detector A vacuum diode UV detector that conve
Teledyne API - T100H Addendum to T100 Operation Manual Theory Of Operation 41 FLOW CONTROL & PRESSURE SENSOR VACUUM PRESSURE SENSORSAMPLE PRESS
Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 42 9.2.2.3. Sample Flow Calculation Unlike the T100, which uses a the
Teledyne API - T100H Addendum to T100 Operation Manual Theory Of Operation 43 9.3. Electronic Operation The following figures replace Figures 10-10
Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 44 AC POWER ENTRANCE ON/OFF SWITCH Sample Gas Pressure Sensor Motherb
45 10. TROUBLESHOOTING & REPAIR For the most part the information contained in Section 11 of the T100 Manual (P/N 06807) is also applicable
TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 46 10.3. Repair Procedures 10.3.1. Repairing the Sample Gas F
47 10.3.2. Sensor Module Repair & Cleaning NOTE: After any repair or service has been performed on the sensor module, the T100H should be a
TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 48 10.3.2.1. Adjusting the UV Lamp (Peaking the Lamp) There a
Teledyne API - T100H Addendum to T100 Operation Manual TROUBLESHOOTING & REPAIR 49 Table 10-3: Example of HVPS Power Supply Outputs UVLAMP_SI
v WARRANTY WARRANTY POLICY (02024D) Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure occur, T-API a
TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 50 10.3.2.2. PMT Hardware Calibration (FACTORY CAL) The senso
Teledyne API - T100H Addendum to T100 Operation Manual TROUBLESHOOTING & REPAIR 51 8. Set the HVPS coarse adjustment to its minimum setting (0
TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 52 20. Adjust gain adjustment potentiometer until the NORM P
Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-1 APPENDIX A - Version Specific Software Doc
APPENDIX A - Version Specific Software Documentation(05244B DCN6038) Model 100EH Instruction Manual A-2 07265A DCN6038
Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-3 APPENDIX A-1: Software Menu Trees, Rev C.0
APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-4 LOW HIGH ZERO SPAN CONC SAMPLE <TST
Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-5 CFG MODEL NAME SERIAL NUMBER SOFTWA
APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-6 NAME EVENT PARAMETERS REPORT PERIOD NUM
Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-7 QUIET COMPUTER SECURITY HESSEN PROTOCOL
Teledyne API - T100H Addendum to T100 Operation Manual vi This page intentionally left blank. 07265A DCN6038
APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-8 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5
Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-9 . YES NO PMT DET WARNINGUV LAMP WARNI
APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-10 5) SAMPLE LED 6) CAL LED 7) FAULT LED
APPENDIX B - Spare Parts Note Use of replacement parts other than those supplied by Teledyne Advanced Pollution Instrumentation (TAPI) may result
This page intentionally left blank. B-207265A DCN6038
069280000 LIST, SPARE PARTS, T100H 3/24/2011 11:04:39 AM PARTNUMBER DESCRIPTION000940400 CD, ORIFICE, .004 BLUE
068810000 PCA, LVDS TRANSMITTER BOARD 069340100 DOM, w/SOFTWARE, STD, T100H * 069500000 PCA, SERIAL & VIDEO INTERFACE
Recommended Spare Parts Stocking Levels (Reference 04527 24Mar2011)Part Number Description1 2-5 6-10 11-20 21-30 UNITS000940800 CD, OR
Part NumberDescription Quantity009690100 KIT, TFE FILTER ELEMENTS, 47MM, 5UM (30) 1018080000 AKIT, DESSICANT BAGGIES (12) 1FL0000001 FILTER, SS2HW0000
Appendix C Warranty/Repair Questionnaire Models T100H, 100EH (05029D DCN 6038) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected]
vii ABOUT THIS ADDENDUM This T100H addendum, PN 07265, is to be used in conjunction with the T100 manual, PN 06807; it is comprised of multiple
Appendix C Warranty/Repair Questionnaire Models T100H, 100EH (05029D DCN 6038) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected]
viii This page intentionally left blank. 07265A DCN6038
ix TABLE OF CONTENTS 1. INTRODUCTION...........................