Teledyne API T802 Paramagnetic O
2
Analyzer Operation Manual
35
Figure 3-8: Current Loop Option Installed
CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD
VOLTAGE OUTPUTS
CAUTION
Servicing or handling of circuit components requires electrostatic discharge
(ESD) protection, i.e. ESD grounding straps, mats and containers. Failure to
use ESD protection when working with electronic assemblies will void the
instrument warranty. See Section 0 for more information on preventing ESD
damage.
To convert an output configured for current loop operation to the standard 0 to 5
VDC output operation:
1. Turn off power to the analyzer.
2. If a recording device was connected to the output being modified, disconnect
it.
3. Remove the top cover as follows:
07275B DCN6418