Analyzer Operation Manual
153
7.1.4.7. RS-232 REPORT FUNCTION
The DAS can automatically report data to the communications ports, where they
can be captured with a terminal emulation program or simply viewed by the user
using the APICOM software.
To enable automatic COM port reporting, follow the instruction shown in
Section 7.1.4 then press:
Starting at the EDIT CHANNEL MENU
SETUP X.X 0) CONC: ATIMER 2, 4032, RS232
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME: CONC
SET> EDIT EXIT
Continue pressing <SET or SET> until ...
SETUP X.X RS-232 REPORT:ON
<SET SET> EDIT PRNT EXIT
Use the PREV and
NEXT buttons to
scroll to the DATA
CHANNEL to be
edited
SETUP X.X RS-232 REPORT: ON
OFF ENTR EXIT
EXIT discards the new
setting
ENTR accepts the
new setting
Toggle to turn the
RS-232 REPORT
feature
ON/OFF
07275B DCN6418